Features-SP365UV能量计系列
管理员 发布日期:2015-5-28 点击数:4133次
产品简介:As UV spot curing technology has evolved, it has become increasingly important to establish a method of measuring system performance.
产品介绍:
Features:·Measures UV irradiance up to 19.99W/cm2(standard range
version) or 99.9W/cm2(extended range version).
·Compact size, portable
·Accepts multiple light guide sizes
·Totally self-contained, battery-powered
Applications:
·Monitor UV spot curing system performance
·Measure individual UV lamp performance
·Measure light guide degradation
·Optimize light guide positioning
·Determine proper UV irradiance required for curing any
material
·Meet ISO-9000 requirements
Introduction:
As UV spot curing technology has evolved, it has become increasingly important to establish a method of measuring system performance. The EIT SpotCure UV Intensity Meter is an easy, portable, effective method of quantifying UV output.
The EIT SpotCure UV Intensity Meter is a self-contained, electro-optic instrument designed to measure and display the irradiance emitted by a UV spot curing system. Degradation of UV lamps, light guides, and reflectors can cause decreases in irradiance and create curing problems. The EIT SpotCure UV Intensity meter is designed to provide the operator with instant feedback as to the performance of the spot curing system.
Configuration:
The EIT SpotCure UV Intensity Meter is totally portable, self-contained, and battery-operated using a special lithium battery stick for extremely long life in excess of 100,000 readings! The measurement head which contains the optics is attached to one end of the cylindrical instrument. Light guide adapters which fit into the measurement head are available to fit most light guides. An EIT SpotCure UV Intensity Meter can be configured to read one of the following UV transmission bandwidths: 250-260nm, 280-320nm, 320-390nm, or 395-445nm.
Operation:
The portable, battery-operated EIT SpotCure UV Intensity Meter is simple to use. Its compact, flashlight-like shape (6.40" long by 1.74" diameter) can comfortably be gripped in one hand.
In operation, the instrument is gripped in one hand while the light guide is inserted with the other hand. Once a spot curing system's light wand is inserted into the measurement head, the "START" switch is depressed and a measurement is taken. When the "START" switch is released, the measurement is frozen and can be viewed on the LCD display. It will be held for approximately 3 minutes until the display times out or until the "START" switch is depressed again. Adaptors insert into the measurement head to accommodate the standard light wands on the market. This flexibility allows the instrument to be used in a variety of monitoring applications.
Short, simple operation instructions are printed on the outside of the instrument. The EIT SpotCure UV Intensity Meter is designed to withstand the rugged UV environment and extremely high intensities which can be associated with UV spot curing. It can monitor UV intensities up to 19.99W/cm2 for the Standard Range Version and 99.9W/cm2 for the Extended Range Version.
SPECIFICATIONS:
Irradiance Range Standard Range Version:
0-19.99W/cm2
Extended Range Version:
0-99.9W/cm2
Resolution Standard Range Version:
10mW/cm2
Extended Range Version:
100mW/cm2
Accuracy +/-5% typical, +/-10% maximum
Display 3 1/2 digit LCD
Power Source Lithium Battery Stick
Battery Life 12,500 hours continuous operation
(over 100,000 readings)
Overall Dimensions 6.40"Lx1.74"Diameter(16.26cm x 4.42cm);
Measurement Head-2.13"Diameter (5.41cm)
Weight 12.8 oz. (358 grams)
Operating Temperature 0-70°C
Display Time 2.5 to 5 minutes
Spectral Ranges Choose
SP254: 250-260nm (UVC);
SP313: 280-320nm (UVB);
SP365: 320-390nm (UVA);
SP405: 395-445nm (UVV)
Case Materials Aluminum, polyester, quartz